L. Gao, BX. Yu, YY. Ding, L. Zhou, LJ. Wen, YG. Xie, ZG. Wang, X. Cai, XL. Sun, J. Fang, Z. Xue, AW. Zhang, QW. LÜ, LJ. Sun, YS. Ge, YB. Liu, SL. Niu, T. Hu, J. Cao, JG. LÜ
The attenuation length measuring device was constructed by using oscilloscope and LabVIEW for signal acquisition and processing. The performance of the device has been tested with a variety of ways, the test results show that the set-up has a good stability and high precision (sigma/mean reached 0.4 percent). Besides, the accuracy of the measurement system will decrease by about 17 percent if a filter is used. The attenuation length of gadolinium-loaded liquid scintillator (Gd-LS) was measured as 15.10 plus or minus 0.35 m where Gd-LS was heavily used in Daya Bay Neutrino Experiment. In addition, one method based on the Beer-Lambert law was proposed to investigate the reliability of the measurement device, the R-square reached 0.9995. Moreover, three purification methods for Linear Alkyl Benzene (LAB) production were compared in the experiment.